Lubefree (LF) replacement turbine for the Midwest Tradition power lever and push button dental handpieces. Stainless steel, lubefree bearings. End cap is NOT included.
Made in the USA. Turbine has a 6 month warranty.
Replacement Parts:
- End cap wrench (#15-97-W)
- Spindle and impeller (#25-303)
- Turbine o-rings (#15-82-OR)
- *For other parts available, please call.
Instructions for Midwest Tradition LF Turbine (PDF):
- Before replacing turbine, carefully examine handpiece head for any dents. If there are any dents, then call American Dental Accessories for repair.
- Then using specialized turbine wrench to unscrew end cap and remove worn turbine out the front of handpiece.
- Inspect head for foreign debris and remove if present. Use a hand scaling instrument to remove any old turbine parts, including any shims, o-rings, washers, or bearing retainers.
- Once all debris and parts have been removed, use a cotton swab and handpiece lubricant to clean handpiece head. Using air from an air/water syringe, blow out handpiece to remove any cotton fibers.
- Turbine includes o-rings, which replace the existing turbine o-rings. One o-ring fits onto the lip on the front bearing of the rubine and rests against a lip inside the head. Second o-ring will fit into the groove in the end cap.
- Holding handpiece face down, insert new turbine fully into the head. Use appropriate tool to screw end cap down until flush against handpiece.
- Before opperating handpiece, install bur securely into turbine. If there is any resistance, then you need to remove the rubine and reinstall again.
- Lastly, install handpiece onto tubing connector and operate foot control to make sure it runs smoothly.