Watch our installation video on how to install your Star 430 push button turbine on your handpiece in minutes! This video show you just how easy it can be.

Star 430 Turbine

Handpiece Turbine Install Instructions:

  1. Using a specialized wrench provided with your high-speed handpiece, unscrew cap from handpiece head.
  2. Remove worn Star 430 turbine, by pushing out from the front.
  3. Use appropriate tool to remove the end cap.
  4. Clean interior of head of foreign material and debris.
  5. Use scaler or explorer instrument to remove old turbine components (ie: shims, o-rings, washers, or bearing retainers).
  6. Take a cotton swab and lubricate handpiece, so you have a clean area.
  7. Use the air from your air/water syringe to blow out handpiece head, so you remove any cotton fibers placed during cleaning.
  8. Then generously lubricate turbine.
  9. Hold handpiece face down and then carefully insert the Star 430 push button turbine into the handpiece head.
  10. Once the turbine is seated, use the appropriate tool to screw the end cap down, so it is flush against the handpiece head.
  11. Secure the end cap tightly with a wrench, so the handpiece turbine is in a safe positioning. If you are unable to fully seat the end cap, remove and check the end cap for debris/old turbine components.

For more handpiece installation information, see our Dental Handpiece Turbine Installation video. Can’t see the video on the installation of a Star 430 turbine? You can also view this video on Youtube.